LED Indicators on the doseBadge5

Modified on Tue, 28 Nov 2023 at 11:42 AM

The doseBadge5 uses a set of LED indicators to show the status of measurements, charging, calibration and general operation.


Four LEDs (orange, green, blue and red) are used in combination to show the current status and activity of the instrument as follows.


Where an LED is flashing, this is shown as Flash Time/Period. For example, 0.1/2.0 is one short flash of 0.1 second every 2 seconds and 0.1/5.0 is one short flash of 0.1 second every 5 seconds.


doseBadge5 on the Charging Dock


StatusLeft LEDRight LEDNotes
Badge charging    
Solid orange    


Badge fully charged    
Solid green    


Badge charging and measuring    
Solid orange    
Flashing blue (0.1/2.0)    

Badge charged and measuring    
Solid green    
Flashing blue (0.1/2.0)    

Badge charging and memory full (or badge error)    
Solid orange    
Solid red    

Badge charged and memory full (or badge error)    
Solid green    
Solid red    

Calibration successful    

Solid blue (5 seconds)    

The badge will revert to the previous indication after 5 seconds

Calibration failed    
Solid red    

The badge will revert to the previous indication after 5 seconds

Instrument start-up    

Solid orange    

Approximately 1 second

doseBadge5 undocked


Status    
Left LED    
Right LED    Notes
Badge on (not measuring)    
Flashing green (0.1/2.0)    


Badge measuring    

Flashing blue (0.1/2.0)    

Badge measuring & dose threshold exceeded    

Flashing red (0.1/2.0)    

Badge paused    

Flashing blue (0.1/5.0)    

Badge paused & dose threshold

exceeded


Flashing red (0.1/5.0)    

Battery low (not measuring)    
Flashing orange (0.1/2.0)    


Battery low & measuring    
Flashing orange (0.1/2.0)    
Flashing blue (0.1/2.0)    

Alternating flashes

Battery low & paused    
Flashing orange (0.1/5.0)    
Flashing blue (0.1/5.0)    

Alternating flashes

Battery low, measuring & dose

threshold exceeded

Flashing orange (0.1/2.0)    
Flashing red (0.1/2.0)    

Alternating flashes

Battery low, paused & dose

threshold exceeded

Flashing orange (0.1/5.0)    
Flashing red (0.1/5.0)    

Alternating flashes

Calibration successful    

Solid blue (5 seconds)    

The badge will revert to the

previous indication after 5 seconds

Calibration failed    

Solid red (5 seconds)    

The badge will revert to the

previous indication after 5 seconds

Instrument start-up    

Solid orange    

Approximately 1 second

Error conditions or warning indications


Status    
Left LED    
Right LED    
Notes
Memory full    

Solid red    

Connect to NoiseTools to download & clear measurements

Badge error    
Solid red    

Connect to NoiseTools to confirm status